Cmm – Kronos

Cmm
  • The KRONOS series is also available in NT version, equipped with the innovative Silicon Carbide spindle for even higher metrological performance.
  • Aluminum alloy moving carriage structure with integrated granite top
  • Generously sized Z spindle made of light aluminum alloy or Silicon Carbide (NT)
  • Mixed rack-and-pinion and hysteresis-free clutch drive system for smooth and repeatable movement
  • Measurement system with high-resolution linear transducers (0.1 µm)
  • Multi-sensor automatic temperature compensation system, with part temperature probe

KRONOS NT 20.15

Each COORD3 CMM is tested according to UNI EN ISO 10360-2 standards.

  • The indicated performance is valid under the following conditions:
  • – PH10M/PH10MQ/PH20/TP20/TP200: Ø4mm stylus length 10mm
  • – PH10MQ/SP25: SM1, stylus Ø 5mm x 50mm
  • – L= measurement length in mm
  • – Ambient temperature conditions:
  • T1 : 18÷22°C; Max gradients: 1.0 °C/h – 2.0 °C/24h – 0.5 °C/m
  • T2: 16÷26°C; Gradients max: 1.0 °C/h – 5.0 °C/24h – 1.0 °C/m
  • [1] Maximum Permissible Error on length measurements according to ISO 10360-2
  • [2] Maximum Permissible Error on Point Grip according to ISO 10360-2
  • [3] Maximum Permissible Error of Point Grip on scanning according to ISO 10360-4 for SP25M/SP80 probes.

KRONOS NT 13.10 – 15.13

  • The indicated performance is valid under the following conditions:
  • – PH10M/PH10MQ/PH20/TP20/TP200: Ø4mm stylus length 10mm
  • – PH10MQ/SP25: SM1, stylus Ø 5mm x 50mm
  • – L= measurement length in mm
  • – Ambient temperature conditions:
  • T1 : 18÷22°C; Max gradients: 1.0 °C/h – 2.0 °C/24h – 0.5 °C/m
  • T2: 16÷26°C; Gradients max: 1.0 °C/h – 5.0 °C/24h – 1.0 °C/m
  • [1] Maximum Permissible Error on length measurements according to ISO 10360-2
  • [2] Maximum Permissible Error on Point Grip according to ISO 10360-2
  • [3] Maximum Permissible Error of Point Grip on scanning according to ISO 10360-4 for SP25M/SP80 probes.

  • Manual Heads: MIH, MH20, MH20i
  • Indexable Motorized Heads: PH10T, PH10M, PH10MQ
  • Point-to-point touchers: TP2, TP20, TP200
  • Analog keyers: SP25 scanning analog keyer
  • Laser sensors: Metris LC/XC series (including qualification sphere)
  • Probe and tip changing systems: Automatic changing stations

  • Temperature range for metrological specifications:
    – T1:
    Ambient temperature: 18 ÷ 22 °C
    Maximum hourly gradient: 1.0 °C/h
    Maximum daily gradient: 2.0 °C/24h
    Maximum gradient in volume: 0.5 °C/m
    – T2:
    Room temperature: 16 ÷ 26 °C
    Maximum hourly gradient: 1.0 °C/h
    Maximum daily gradient: 5.0 °C/24h
    Maximum gradient in volume: 1.0 °C/m
  • Operating Temperature: 15 to 35 °C
  • Relative humidity for operation: 40 to 80 % (non-condensing)
  • Allowable vibration: (Peak/peak acceleration)
    30 mm/s2between 1 and 10 Hz
    15 mm/s2between 10 and 20 Hz
    50 mm/s2between 20 and 100 Hz

  • Hava basıncı: 6 bar (79 PSI)
  • Hava tüketimi : 160 NL/min
  • Voltaj : Single Phaze, 230 V 50 Hz ± 2% single phase
  • Maksimum güç tüketimi : 10A, 1600 W

  • Passive Antivibration System
  • Active vibration damping system (AVM)
  • Multi-wire wiring
  • Manual/automatic pallet loading/unloading system
  • PC and Printer
  • Training at Coord3 center or agents
  • Installation by Coord3 staff or agents
  • Models and dimensions:
    COORD3 KRONOS CMMs are available with the following measuring strokes (Y.Z mm):
    13.10/15.13/20.15